A450C+ FO PL
Temperature Range: 600°C to 3200°C
Digital Fiber Optic IR Pyrometer in Two Color with Mono Fiber Optic Cable
A450C+ FO PL is a highly accurate digital two color pyrometer to provide high performance and low maintenance of non contact temperature measurement in demanding industrial and R&D environments. The
Pyrometer measures the temperature of an object by calculating the ratio of the energies at two different wavelength bands. The ratio technique eliminates and reduces errors in temperature measurement caused by
changes in Emissivity, Surface finish and energy absorbing materials such as Water vapour etc.
The IR pyrometer parameters can be selected via the keypad on the back panel. The settings are displayed on the OLED Display.
Key Features
- Wide temperature ranges from 600°C to 3200°C
- Fast Response Time
- Spectral range 0.7 μm….1.15 μm
- Integrated OLED Display & Parameterizing Keys
- Laser targeting light for high precision targeting
- Two color model (Switchable to single color mode)
- Analog output options 0…20mA, 4…20mA
- RS-485 Serial interface
- User friendly PC software for communication
- Green LED for pyrometer status indication and Red LED for Laser ON/OFF indication (PL)
Scope of Supply
- Pyrometer with OLED Display & Parameterizing Keys
- 2.5 Mtr mono fibre optic cable
- Optical head OH (l), OH(ll), OH(ll)variable or OH(lll) variable
- Laser pilot light(PL)
- Analog output 4…20mA, 0..20mA
- Digital Interface RS-485
- 5 mtr long connection cable with connector
- Calibration certificate, Software & Operation manual
Optional Accessories
- Mechanical & Electrical Accessories
- Extra Length of Fiber Cable and Connection Cable
Technical Specifications
Model | A450C+ FOPL |
Temperature Range (Analog sub-range adjustable) | 600°C to 1600°C 800°C to 2500°C 1000°C to 3200°C |
Spectral Range | 0.7 …. 1.15 μm |
Photodetector Type | Si/Si |
Distance to Spot Size Ratio | 100:1 OH(I) 200:1 OH(II) 200:1 OH(ll) – Variable 400:1 OH(lll) – Variable |
Emissivity Slope (ε) | 0.1….1.0 adjustable adjustable (Single color mode) |
Emissivity Slope (ε1/ε2) | 0.75….1.25 slop adjustable (Two color mode) |
Response Time | 20 msec adjustable upto 10 sec |
Accuracy | ± 0.5% of the measured value + 1°C |
Repeatability | 0.1% of reading in °C + 1°C |
Sighting Option | Laser Pilot Light(PL) |
Analog Output | 0-20mA, 4-20mA (User selectable) |
Digital Output | RS-485 |
Operating Temp. Range | 0°C…….70°C Optical Head and Fiber Optic Cable upto 250°C |
Storage Temp. Range | -20°C…70°C |
Adjustable Parameters, Features via Software and Android app | Emissivity, Response Time, Clear Time (Peak Picker), Analog Output, Analog Scale (Sub range), Unit Of Temperature(°C/°F), Communication mode (Comm. mode), Record feature, Relay set point & hysteresis, etc. |
Adjustable Parameters, Features via Keypad | Emissivity, Response Time, Clear Time (Peak Picker), Analog Output, Analog Scale (Sub range), Unit Of Temperature(°C/°F), Communication mode (Comm. mode), Relay set point & hysteresis, address, etc |
Relay output | One Relay output with hysteresis 60V DC / 42 AC RMS, 0.4 A |
Power Supply | 12V to 28V DC with reverse voltage protection |
Power Consumption | Max 4.0 watt |
Laser Power | <1 m watt (only PL model) |
Protection Class | Ip65 |
Housing | Stainless Steel |
Isolation | Power Supply, Digital Output and Analog Output are galvanically islolated against each other |
Operating Humidity | 10-95%, Non-Condensing Conditions |
Weight & Dimensions | 1200g Dia= Ø 56mm; Length=188.5mm |