AL45

Temperature Range: 400°C to 1500°C

Highly Accurate Non-Contact Pyrometer With Analogue Output and Digital Interface

AL45 are special pyrometers which provides the advantage of measurement of flames and combustion gases that include Co2. This format combines the high accuracy of digital signal processing with the simple connection.

AL45 IR Pyrometer is used for target temperature ranging from 400°C to 1500°C. The electronic assembly is protected by IP65 rugged stainless steel housing. AL45 IR Pyrometer is provided with USB 2.0 for parameter setting & Data Logging.

Category:

Key Features

  • Measurement of flames and combustion gases that include CO2.
  • Highly Accurate due to digital linearization of the output.
  • Temperature range from 400°C to 1500°C
  • Spectral range of about 4.43 µm (absorption band for hot CO )2
  • Four wire form with Analog output 4…20mA or 0…20mA or 0…10V
  • Serial communication RS-232/RS-485/RTU Modbus Serial Interface (User Selectable)
  • Response time 60 ms
  • Very good stability
  • Laser Targeting for precision targeting
  • User friendly Software

Scope of Supply

  • Pyrometer with Laser pilot light(PL)
  • Analog output 4…20mA,0..20mA, 0..10V
  • USB 2.0 Digital output, USB Cable
  • RS-232/RS-485 /RTU Modbus Serial Interface (User Selectable)
  • 5 mtr long connection cable with connector
  • 1.5 mtr. Communication Cable
  • Calibration certificate, Software & Operation manual

Optional Accessories

  • Mechanical and Electrical Accessories
  • Extra Cable Lengths

Models at a Glance

ModelAL45
Temperature Range
(Analog sub-range adjustable)
400°C to 1500°C
Spectral Rangeabout 4.43 µm (absorption band for hot CO2
Photodetector TypeThermopile
Distance to Spot Size Ratio40:1
Emissivity (ε)0.1….1.2 adjustable
Response Time60 msec adjustable upto 10sec
AccuracyT < 500°C; ± 1.5% of measured value
T ≥ 500°C; ± 1% of measured value
(The instrument must be at constant ambient temperature for minimum 25-30 mins in
power on condition)
Repeatability0.3% of reading in °C +1°C
Sighting OptionLaser Pilot Light (PL)
Analog Output0-20mA, 4-20mA, 0-10V (User selectable)
Digital OutputUSB 2.0
RS-232/RS-485/RTU Modbus Serial Interface (User selectable)
*At a time only one digital output possible
Operating Temp. Range0°C…….70°C
0°C…..200°C (With water cooling jacket)
Storage Temp. Range-20°C…70°C
Adjustable Parameters, Features via Software Emissivity, Response Time, Clear Time(Peak Picker), Analog Output, Analog Scale (Sub
Range), Unit Of Temperature(°C/°F), Communication mode (Comm.mode), Record
feature etc
Power Supply24V DC
Power ConsumptionMax 2.5 watt
Laser Power<1 m watt
Protection ClassIp65
HousingStainless Steel
IsolationPower supply,*Digital output and Analog output are galvanically islolated against each
other
* Not applicable for USB 2.0 digital output
Operating Humidity10-95%, Non-Condensing Conditions
Weight & Dimensions600g, Dia= Ø 49.5 mm; Length=118mm