LBBCH SP

Temperature range (Standard)*: -25°C to 100°C

Extended area black body is defined by the large emitting surface area precise temperature control with good uniformity. Tempsens make Blackbodies are state of the art, highly accurate and stable with different standard sizes and temperature ranges. The LBBCH SP Series Extended Area black bodies are low temperature infrared reference sources operating either in absolute or differential mode.

Specifications

ParameterLBBCH SP
Emissive area50 x 50 mm²
Temperature range (Standard)*-25°C to 100°C
Emissive area uniformity(1)±0.2°C @ 50°C or better
Emissivity0.98±0.02
Stability ±0.02°C or better
Heating Time15 Min
Cooling Time20 Min
Display resolution0.01°C
Method of controlDigital self tuned PID Controller
Head dimensions W x H x D (mm³)300(W) X 320(H) X 190(D) mm
Weight15 kg
Max. power consumption500 W
Power supply230 VAC, 1 ph. 50 Hz
Operating temperature range (head)10°C to +25°C

* At 23°C Ambient
*1 at 80% of emissive area

Key Features

  1. Wide Temperature Range
  2. Large emissive area
  3. High Emissivity
  4. Accuracy and performance
  5. Easy to use
  6. Computer Interface

Accessories

  • Master Pyrometer (Optional)
  • Calsoft Software
  • Operational Manual
  • Carry Case (Only for LBB11CH Model)
  • NABL accredited calibration certificate (Optional)

Standards

  • Wide Temperature Range
  • Large emissive area
  • High Emissivity
  • Accuracy and performance
  • Easy to use
  • Computer Interface