T3-814
Temperature Range: -20°C to 1000°C
Highly Accurate On-Line Infrared Non-Contact Pyrometer in Two Wire Technology
T3-814 PL Digital Infrared Pyrometers with Thermopile Sensors provides the advantage of noncontact temperature measurement of non-metallic surfaces, painted, coated or anodized metals etc. T3-814 PL IR Pyrometer is designed for easy integration into standard 2 wire system. This format combines the high accuracy of digital signal processing with the simple connection. T3-814 PL Infrared Pyrometer is used for target temperature ranging from -20°C to 1000°C. The electronic assembly is protected by IP 65 rugged stainless steel housing.
The parameters like Emissivity can be adjusted directly at device. Other parameters like Analog Sub Range, Response Time, Peak Picker, Valley Picker etc can be preset ex-works or adjustable via Software Infrasoft. T3 Series is provided with TTL output. Using the USB interface Infralink and Software pyrometer can be connected to PC for parameter setting and data logging.
Key Features
- Two wire technology for easy electrical connection
- Wide temperature range -20°C to 1000°C
- Spectral range 8.0 μm….14 μm
- Laser Light for precison targeting of the measuring object
- 60 ms response time adjustable upto 10 sec
- Analog output 2 wire….4-20 mA (Isolated)
- Digital TTL output & Infrasoft software
- Small spot sizes
- Emissivity can be adjusted directly at device
- Accessories for mounting and cooling options
Scope of Supply
- Pyrometer with 3 mtr. long connection cable
- Analog output 2 wire….4-20 mA
- Emissivity adjustable switch
- TTL output
- Calibration certificate, Operation manual
Optional Accessories
- Extra cable lengths
- Mechanical and Electrical Accessories
- USB Interface card & PC Software
Technical Specifications
Model | T3-814 |
Temperature Range | -20°C to 1000°C 75°C to 1000°C |
Spectral Range | 8 μm…14 μm |
Photodetector Type | Thermopile |
Distance to Spot Size Ratio | 50:1 100:1 |
Emissivity (ε) | 0.1….1.0 adjustable |
Response Time | 60 msec. adjustable upto 10sec |
Accuracy | T < 200°C; ± 1.5% of measured value or 3°C whichever is greater T ≥ 200°C; ± 1% of measured value or 4°C whichever is greater (The instrument must be at constant ambient temperature for minimum 25-30 mins in power on condition |
Repeatability | 0.3% of reading in °C + 1°C |
Sighting Option | Laser Pilot Light (PL) |
Analog Output | 2 wire….4-20mA Linear Temperature Output Load : Max 500Ω at 24V DC, Max 200Ω at 18V DC |
Digital Output | TTL |
Operating Temp. Range | 0°C…….70°C 0°C…..200°C (With water cooling jacket |
Storage Temp. Range | -20°C…70°C |
Adjustable Parameters, Features via Software and Parameterizing Keys | Emissivity, Response Time, Analog Scale (Sub Range), Unit of Temperature (°C/°F), Peak Picker, Valley Picker |
Power Supply | 24 V DC (5 to 25 V DC for Laser Targeting light (I≤30 mA) |
Power Consumption | For Laser Targeting Max 0.65 watt For Device Max 0.6 watt |
Laser Power | <1 m watt |
Protection Class | Ip65 |
Housing | Stainless Steel |
Operating Humidity | 10-95%, Non-Condensing Conditions |
Weight & Dimensions | 250g Dia=Ø40mm; Length=113.5mm |